Company record
Olympus Corporation of the Americas
0 FDA warning letters since 2023 and 125 recalls in the last 12 months.
Get an email the day FDA posts anything new about Olympus Corporation of the Americas.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Oct 7 | Class II | Olympus Corporation of the Americas Device |
Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A. There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert. |
| Oct 7 | Class II | Olympus Corporation of the Americas Device |
Olympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunction wtih OLYSENSE Hub System While in the high-power operating mode, a video dropout/blackout may occur when using a particular accessory cable. |
| Oct 7 | Class II | Olympus Corporation of the Americas Device |
Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A. There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert. |
| Sep 30 | Class II | Olympus Corporation of the Americas Device |
Olympus BiCOAG Hemostasis Probe 7Fr Fixed Pin Connector. Model/Catalog Number: CD-B622LA. Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive. |
| Sep 30 | Class II | Olympus Corporation of the Americas Device |
Olympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA. Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive. |
| Sep 30 | Class II | Olympus Corporation of the Americas Device |
Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Model Number: BD-420X-1355. Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm, |
| Sep 30 | Class II | Olympus Corporation of the Americas Device |
Olympus BiCOAG Hemostasis Probe 7Fr Coaxial Connector. Model/Catalog Number: CD-B620LA. Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive. |
| Sep 30 | Class II | Olympus Corporation of the Americas Device |
Olympus BiCOAG Hemostasis Probe 10Fr Fixed Pin Connector. Model/Catalog Number: CD-B612LA. Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive. |
| Sep 2 | Class II | Olympus Corporation of the Americas Device |
POWERSEAL 5 MM (in diameter) 23 CM (long), Curved Jaw Sealer & Divider, Double Action. GEI / Electrosurgical, cutting & coagulation & accessories Potential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device. |
| Jul 29 | Class II | Olympus Corporation of the Americas Device |
Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4 Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases. |
| Jul 29 | Class II | Olympus Corporation of the Americas Device |
Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color. Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities. |
| Jul 22 | Class II | Olympus Corporation of the Americas Device |
Olympus Ellik Evacuator Includes Glass Body, Male Connecting Tubing & Latex Bulb. Model Number: 194. Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices. |
| Jul 22 | Class II | Olympus Corporation of the Americas Device |
Olympus Toomey Evacuator 30cc with Adaptor. Model Number: 188. Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices. |
| Jul 22 | Class II | Olympus Corporation of the Americas Device |
Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A. Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices. |
| Jul 22 | Class II | Olympus Corporation of the Americas Device |
Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A. Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices. |
| Jul 15 | Class II | Olympus Corporation of the Americas Device |
Olympus Single Use Distal Cover, Catalog Number MAJ-2315, Used with: EVIS EXERA III Duodenovideoscope TJF-Q190V- designed to be attached to Olympus duodenovideoscopes to cover the distal end of the insertion tube and aro… Olympus has developed and released newly designed version of the Distal Cover (MAJ-2315) to replace previous design of the Single Use Distal Covers. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 45cm. Model Number: TB2-0545FC. Potential for detachment of a distal tip component of the device during use. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115. XXX |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 20cm. Model Number: TB2-0520FC. Potential for detachment of a distal tip component of the device during use. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2113. Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2119. Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2114. Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2138. Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2118. Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 35cm. Model Number: TB2-0535FC. Potential for detachment of a distal tip component of the device during use. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2112 Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117. Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2110. May be included with the Olympus OER-ELITE. Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 25cm. Model Number: TB2-0525FC. Potential for detachment of a distal tip component of the device during use. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2111 Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2116. Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 29 | Class II | Olympus Corporation of the Americas Device |
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2330. Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely. |
| Apr 8 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device asse… Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing. |
| Apr 8 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly whi… Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulatio… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443. The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. C… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140. The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate and apply radio-frequency (RF) alterna… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Olympus Endoscope Reprocessor OER-Elite. Model Number: N5789340. The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Olympus Endoscope Air/Water Valve - MAJ-1444. Model Number: MAJ-1444. The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus Everest Bipolar Cutting Forceps, 5 mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3005 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are inten… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulat… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm l… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. |
| Mar 11 | Class II | Olympus Corporation of the Americas Device |
Olympus Single use Biopsy Valve. Model/Catalog Number: MAJ-210. 20 units per box. The MAJ-210 has been designed to be attached to the instrument channel port of compatible endoscopes to prevent reflux of body fluids. Potential for rubber fragment detachment during use. |
| Mar 11 | Class II | Olympus Corporation of the Americas Device |
Olympus Single Use Biopsy Valve. Model/Catalog Number: MAJ-1218. 20 units per box. The MAJ-1218 has been designed to be attached to the instrument channel port of compatible endoscopes to prevent leakage of body fluid… Potential for rubber fragment detachment during use. |
| Mar 11 | Class II | Olympus Corporation of the Americas Device |
Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Catalog Number: SPL-G Product Description: The ShockPulse Lithotripsy System is an electromechanical device… Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues. |
| Mar 11 | Class II | Olympus Corporation of the Americas Device |
Brand Name: ShockPulse-SE Lithotripsy System SPL-SR Product Name: ShockPulse-SE Lithotripsy System - Reusable Probes Model/Catalog Number: SPL-SR Product Description: The ShockPulse Lithotripsy System is an electromec… Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues. |
| Mar 11 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An asse… Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices. |
| Mar 11 | Class II | Olympus Corporation of the Americas Device |
Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPulse Lithotripsy System is an electromec… Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Sheath, 10 Fr. Model No. A37004A. for endoscopic diagnosis and treatment in urological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath, 24 Fr., with Deflecting Obturator; Model No. A22041T. Resection sheath for urologic applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Inner Sheath, Long; Model No. WA22017A. Inner sheath for urological and gynecological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath, 28 Fr.; Model No. A22043A. Resection sheath for urologic applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Inner Sheath. Model No. A2641. for endoscopic diagnosis and treatment in urological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath, 28 Fr. with Deflecting Obturator. Model No. A22043T. Resection sheath for urologic applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Inner Sheath; Model No. A2660. for endoscopic diagnosis and treatment in urological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Inner Sheath, for 26 Fr. Outer Sheath. Model No. A22040A. Resection sheath for urologic applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath, 26 Fr. with Deflecting Obturator; Model No. A22042T. Resection sheath for urologic applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath, 8 mm. Model No. A42011A. Resection sheath for gynecological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath; Model No. A2666. Resection sheath for urologic applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Inner Sheath. Model No. A4741. for endoscopic diagnosis and treatment in gynecological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Inner Sheath for 26 Fr. Outer Sheath. Model No. A22040T. Resection sheath for urologic applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath. Model No. A2666T. for endoscopic diagnosis and treatment in urological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Inner Sheath 21 Fr. Model No. A2660T. for endoscopic diagnosis and treatment in urological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath, 24 Fr; Model No. A22041A. Resection sheath for urologic applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath, 26 Fr.; Model No. A22042A. Resection sheath for urologic applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Cystoscope Outer Sheath; Model/Catalog number: WA22810A; Product being removed due to incompatibility when used in conjunction with a GreenLight Laser for BPH therapy which could damage the tip of the device. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Resection Sheath. Model No. A22014A and A22014T. for endoscopic diagnosis and treatment in urological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Inner Sheath. Model No. A2642. for endoscopic diagnosis and treatment in urological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Mar 4 | Class II | Olympus Corporation of the Americas Device |
Olympus Inner Sheath, Long; Model No. WA22017T. Inner sheath for urological and gynecological applications. Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat 5 mm, 45 cm, Front-Actuated Grip Type S Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat, 5mm, 35 cm, Pistol Grip Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat 5 mm, 20 cm, Front-Actuated Grip Type S Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class I | Olympus Corporation of the Americas Device |
Olympus High Flow Insufflation Unit. Model Number: UHI-3. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdo… Issue with software algorithm which may lead to overpressure events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat, 5 mm, 20 cm, Inline Grip Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class I | Olympus Corporation of the Americas Device |
Olympus High Flow Insufflation Unit. Model/ Number: UHI-2. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abd… Issue with software algorithm which may lead to overpressure events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat 5 mm, 35 cm, Front-Actuated Grip Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat, 5 mm, 45 cm, Pistol Grip Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat 5 mm, 10 cm, Inline Grip Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class I | Olympus Corporation of the Americas Device |
Olympus High Flow Insufflation Unit. Model Number: UHI. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdomi… Issue with software algorithm which may lead to overpressure events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat, 5 mm, 35 cm, Front-Actuated Grip Type S Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat, 5 mm, 45 cm Front-Actuated Grip Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat, 5 mm, 35 cm, Inline Grip Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat, 5 mm, 45 cm Inline Grip Firm is initiating a removal due to continued reports of adverse events. |
| Feb 25 | Class II | Olympus Corporation of the Americas Device |
Olympus Thunderbeat 5 mm, 20 cm, Front-Actuated Grip Firm is initiating a removal due to continued reports of adverse events. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using h… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0330; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewir… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0725; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewir… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3025 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using h… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-431Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0725; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus e… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus e… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewir… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus … Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Brand Name: CleverCut Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0320; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus … Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number:KD-VC433Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus en… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewir… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0320; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus e… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Single Use Electrosurgical Knife Product Name: Olympus Single Use Electrosurgical Knife KD-640L Model/Catalog Number: KD640-L Product Description: An electrosurgical device consisting of a handpiece with m… A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus e… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3030 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using h… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Single Use Electrosurgical Knife KD-645 Product Name: Olympus Single Use Electrosurgical Knife KD-645 Model/Catalog Number: KD-645L Product Description: An electrosurgical device consisting of a handpiece … A deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3020 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using h… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewir… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0330; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus e… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Disposable Triple Lumen Sphincterotome Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0320 Product Description: These instruments have been designed to be used with an O… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC412Q-0215; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus e… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papil… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1530 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using h… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0725 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewir… Devices which did not undergo thermoforming could deform and lose performance. |
| Feb 11 | Class II | Olympus Corporation of the Americas Device |
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC433Q-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus e… Devices which did not undergo thermoforming could deform and lose performance. |
| Dec 3, 2025 | Class I | Olympus Corporation of the Americas Device |
Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Model/Catalog Number: HX-400U-30 Used with an Olympus endoscope to deliver a nylon loop snare. Potential that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally fixed around patient anatomy. |
| Dec 3, 2025 | Class II | Olympus Corporation of the Americas Device |
Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended to be used with an Olympus video system center, light sourc… Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users. |
| Nov 26, 2025 | Class II | Olympus Corporation of the Americas Device |
Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate ne… Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit. |
| Nov 19, 2025 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmen… The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use. |
| Nov 19, 2025 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use Electrosurgical Snare Model/Catalog Number: SD-U400U-15 Product Description: These instruments have been desig… Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton. |
| Nov 19, 2025 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an … Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton. |